Anechoic Chambers for Antenna, Radar and Wireless Testing
An anechoic chamber is not a room with foam on the walls. It is a quiet zone of a stated size, at a stated distance, inside a shielding enclosure whose penetrations are accounted for. If a quote does not name the quiet zone, it has not told you what you are buying.
What we supply
Microwave anechoic chambers for antenna pattern and gain measurement, RCS work, radar and microwave component testing, and wireless device certification. Absorber on all six faces, or five faces plus a reflective floor where a ground plane is required for EMC work.
The HM-MC antenna chamber, in numbers
| Item | Value |
|---|---|
| Typical outer dimensions | 7.0 × 4.0 × 3.0 m (L × W × H), other sizes on request |
| Shield module | 3 m × 1 m, double-sided hot-dip galvanised steel ≥ 2 mm, zinc layer ≥ 20 μm |
| Module joint | conductive gasket between adjacent modules, M10 bolted |
| Absorber | HM-AM pyramidal PU or EPP, sized to the lowest frequency of interest |
| Shielded door | 1.0 × 2.0 / 1.2 × 2.0 / 1.5 × 2.0 m — manual, electric or pneumatic |
| Turntable | ø1.2 m or ø1.5 m, 300 kg or 500 kg load |
| Antenna mast | 0.9 m to 2.0 m working height |
| Raised floor | 300 mm |
| Shielding performance | meets and typically exceeds EN 50147-1 |
The construction order from inside out is: absorber, anti-corrosion plywood backing, shielding module. The plywood is not decoration — it spreads the load of the absorber and stops the pyramid bases from wearing against the steel modules when the room is re-lined years later.
Quiet zone is a function of absorber length, not room size
This is where most small-chamber projects go wrong. A 200 mm pyramid stops absorbing usefully around 300 MHz at normal incidence, so a room lined only with 200 mm pyramid has no quiet zone below that. To reach 30 MHz you need either very long pyramids (which eat metres of interior volume) or hybrid ferrite tiles under the pyramid.
| Absorber | Height | Useful from | Typical use |
|---|---|---|---|
| AM-30 | 30 mm | ≈ 3 GHz | high-frequency only, small test boxes |
| AM-100 | 100 mm | ≈ 1 GHz | compact antenna chambers |
| AM-200 | 200 mm | ≈ 300 MHz | general microwave work |
| AM-500 | 500 mm | ≈ 100 MHz | wide-band antenna and RCS |
| Ferrite 6.7 mm + pyramid | hybrid | ≈ 30 MHz | full EMC band without huge pyramids |
Rule of thumb we use in quoting: interior size must be at least the quiet zone plus two absorber lengths plus one wavelength of clearance. Anything less and the quiet zone is theoretical.
Ferrite tile specification
| Item | Value |
|---|---|
| Tile size | 99.95 × 99.95 mm |
| Thickness | 6.7 mm or 5.5 mm |
| Weight | 0.33 kg (6.7) / 0.28 kg (5.5) |
| Operating range | 30 MHz to 1 GHz, extended to 18 GHz by hybrid pyramid |
| Max service temperature | 85 °C |
| Fire | non-flammable, sintered ferrite base |
| Mounting | screwed to structure or glued to shielding enclosure |
| Manufacturing | all six faces ground flat, to minimise tile-to-tile gaps |
The gap between tiles matters more than the tile itself. Unground tiles stacked by hand leave 1–2 mm gaps, and at 300 MHz that is the difference between a working chamber and one that fails. Ask whether the tiles are surface-ground on all six faces.
Acceptance tests
- Shielding effectiveness to EN 50147-1 / GB/T 12190, typically 14 kHz to 18 GHz or 40 GHz
- Normalised site attenuation (NSA) where the room is used for EMC, to CISPR 16-1-4 and ANSI C63.4
- Field uniformity to IEC 61000-4-3 where immunity testing is in scope
- Site VSWR from 1 GHz up, where the room is used above 1 GHz
- Quiet zone verification — antenna pattern and gain against a reference antenna
Frequently asked
Can the chamber be assembled inside an existing room?
Yes — modular panels are 3 m × 1 m and pass through a standard 1.2 m door. A 7 × 4 × 3 m chamber has been installed in a basement using only a passenger lift and a hand truck.
Who carries out the acceptance test?
An independent accredited laboratory. We do not accept our own chamber with our own instrument; for CNAS, DILAC or military laboratory certification, the third-party report is what the auditor wants.
Can you upgrade an existing chamber?
Usually yes — re-lining absorber, replacing doors and seals, adding filters and vents. Performance of the shell rarely degrades; the doors, gaskets and penetrations usually do.
Where we have shipped
Market notes and project references, by country: