MRI RF Shielded Doors: Fully Non-Magnetic, 90 dB from 10 to 100 MHz
An MRI door is not a small shielded door. It is a door that must attenuate at the Larmor frequency of the magnet, swing without any ferromagnetic part anywhere near the bore, and allow a patient trolley through with a 5 mm gap at the sides. Get any one of those wrong and the scanner will show artifact on every scan.
The MRI door in numbers
| Item | Value |
|---|---|
| Clear entrance | 1200 × 2000 mm (standard medical trolley clearance) |
| Frame size | 1320 × 2200 mm |
| Attenuation | 90 dB from 10 MHz to 100 MHz |
| Body | pre-processed wooden panel, copper-clad on both faces |
| Hardware | purpose-shaped brass fittings, stainless steel screws, beryllium copper finger strip |
| Magnetic content | none — the door will not respond to the magnet fringe field |
| Threshold | direction can be modified during assembly |
| Spares | seal and hardware replaceable on site |
Why 90 dB at 10 to 100 MHz, and not 100 dB
A 1.5 T scanner works at 63.9 MHz; a 3 T scanner at 127.7 MHz. The door must attenuate across the band that covers both, with margin. 90 dB is the figure we test and publish, because it is what a 1.2 m clear opening can hold in service with a copper-clad wooden leaf — the leaf cannot be steel, and copper on wood is a different engineering problem from copper on steel.
Note where the requirement really sits: a scanner room specification usually asks for 100 dB, and the door then becomes the limiting element. Ask your scanner vendor what attenuation the room must hold at the operating frequency, and design the door as the last link in that chain rather than the first.
Non-magnetic, all the way down
The failure mode in MRI doors is almost never the shielding. It is a stainless screw that turned out to be magnetic grade, or a hinge pin, or a latch spring. In a 1.5 T fringe field even small ferrous parts produce artifact, and they also get pulled. Our fittings are brass, made in-house on our own dies, stainless is verified non-magnetic before assembly, and the beryllium copper strip is checked as a batch.
What else a scanner room needs from us
- RF shielded doors — this page, the patient entrance
- Waveguide penetrations — fibre optic, quench pipe, gas lines, patient monitoring cables
- Power filters — non-magnetic enclosure, filtered at the room wall
- Shielded windows — 80 to 300 mesh, for the control room view
- RF shielding panels — copper or galvanised, welded, forming the room itself
We supply the RF enclosure and penetrations. We do not supply the magnet, the gradient coils, or the magnet room's quench system — those come from the scanner vendor, and the interface between their requirements and our shell is where the design meeting has to happen.
Frequently asked
Does the door need to be a different size for a 3 T scanner?
The clearance is the same — a patient trolley is a patient trolley. What changes is the attenuation band and the margin at the higher Larmor frequency. Tell us the field strength and the vendor and we will state the attenuation at the operating frequency.
Can the door be fitted to an existing room?
Yes, if the wall opening matches or can be adapted with a frame. The frame carries the seal, so the opening must be built to the frame, not the other way round.
How is the door tested?
At the operating frequency of the scanner, with the door closed and latched, using the same method the room acceptance uses. We can supply the door test data; the room acceptance should be done by the installing contractor with the scanner vendor present.
Where we have shipped
Market notes and project references, by country:
Ask about MRI RF doors